Continuous Test Methods Updated
June 3, 2009
On May 22, 2008, the
EPA published a final rule in the Federal Register that made
technical corrections to five test methods. Inadvertent printing errors were
made in the publication. Text insertions were misplaced, duplicate insertions
were made, and the definition for system bias was inadvertently revised. On
Friday, the agency re-released the test method changes, with corrections.
The revised version
of the pertinent section (Method 3A, Section 7.1) is below:
PART
60--STANDARDS OF PERFORMANCE FOR NEW STATIONARY SOURCES
Method
3A--Determination of Oxygen and Carbon Dioxide Concentrations in Emissions From
Stationary Sources (Instrumental Analyzer Procedure)
* * * * *
7.1 Calibration Gas.
What calibration gasses do I need? Refer to Section 7.1 of Method 7E for the
calibration gas requirements. Example calibration gas mixtures are listed
below. Precleaned or scrubbed air may be used for the O2
high-calibration gas provided it does not contain other gases that interfere
with the O2 measurement.
(a)
CO2 in nitrogen (N2).
(b)
CO2 in air.
(c)
CO2/SO2 gas mixture in
N2.
(d)
O2/SO2 gas mixture in
N2.
(e)
O2/CO2/SO2
gas mixture in N2.
(f)
CO2/NOX gas mixture in
N2.
(g)
CO2/SO2/NOX
gas mixture in N2.
The tests for
analyzer calibration error and system bias require high-, mid-, and low-level
gases.
* * * * *
Appendix A-4-- [Amended]
3. Method 7E is
amended as follows:
a. By revising
section 3.12.
b. By revising
section 3.16.
c. By revising
section 7.1.
d. By revising
section 8.1.2.
e. By revising
section 8.2.1.
f. By revising
section 8.2.4.
g. By revising the
Summary Table of QA/QC in Section 9.0.
h. By revising
section 12.11.
i. By revising
section 16.2.2.
Method
7E--Determination of Nitrogen Oxides Emissions From Stationary Sources
(Instrumental Analyzer Procedure)
* * * * *
3.12
Low-Concentration Analyzer means any analyzer that operates with a calibration
span of 20 ppm NOX or lower. Each analyzer model used
routinely to measure low NOX concentrations must pass a
manufacturer's stability test (MST). An MST subjects the analyzer to a range of
line voltages and temperatures that reflect potential field conditions to
demonstrate its stability following procedures similar to those provided in 40
CFR 53.23.
Ambient-level
analyzers are exempt from the MST requirements of Section 16.3. A copy of this
information must be included in each test report.
Table 7E-5 lists the
criteria to be met.
* * * * *
3.16 System Bias
means the difference between a calibration gas measured in direct calibration
mode and in system calibration mode.
System bias is
determined before and after each run at the low- and
mid- or
high-concentration levels. For dilution-type systems, pre- and post-run system
calibration error is measured rather than system bias.
* * * * *
7.1 Calibration Gas.
What calibration gases do I need? Your calibration gas must be NO in N2
and certified (or recertified) within an uncertainty of 2.0 percent in
accordance with "EPA Traceability Protocol for Assay and Certification of
Gaseous Calibration Standards" September 1997, as amended August 25, 1999,
EPA-600/R-97/121. Blended gases meeting the Traceability Protocol are allowed
if the additional gas components are shown not to interfere with the analysis.
If a zero gas is used for the low-level gas, it must meet the requirements under
the definition for "zero air material" in 40 CFR 72.2. The
calibration gas must not be used after its expiration date. Except for
applications under part 75 of this chapter, it is acceptable to prepare
calibration gas mixtures from EPA Traceability Protocol gases in accordance
with Method 205 in appendix M to part 51 of this chapter. For part 75
applications, the use of Method 205 is subject to the approval of the
Administrator. The goal and recommendation for selecting calibration gases is
to bracket the sample concentrations. The following calibration gas
concentrations are required:
* * * * *
8.1.2 Determination
of Stratification. Perform a stratification test at each test site to determine
the appropriate number of sample traverse points. If testing for multiple
pollutants or diluents at the same site, a stratification test using only one
pollutant or diluent satisfies this requirement. A stratification test is not
required for small stacks that are less than 4 inches in diameter. To test for
stratification, use a probe of appropriate length to measure the
NOX (or pollutant of interest) concentration at twelve
traverse points located according to Table 1-1 or Table 1-2 of Method 1.
Alternatively, you may measure at three points on a line passing through the
centroidal area. Space the three points at 16.7, 50.0, and
83.3 percent of the
measurement line. Sample for a minimum of twice the system response time (see
Section 8.2.6) at each traverse point.
Calculate the individual
point and mean NOX concentrations.
If the concentration
at each traverse point differs from the mean concentration for all traverse
points by no more than: (a) ±5.0 percent of the mean concentration; or (b) ±0.5
ppm (whichever is less restrictive), the gas stream is considered unstratified
and you may collect samples from a single point that most closely matches the
mean. If the 5.0 percent or 0.5 ppm criterion is not met, but the concentration
at each traverse point differs from the mean concentration for all traverse
points by no more than: (a) ±10.0 percent of the mean; or (b) ±1.0 ppm
(whichever is less restrictive), the gas stream is considered to be minimally
stratified, and you may take samples from three points. Space the three points
at 16.7, 50.0, and 83.3 percent of the measurement line.
Alternatively, if a
twelve-point stratification test was performed and the emissions were shown to
be minimally stratified (all points within ±10.0 percent of their mean or
within ±1.0 ppm), and if the stack diameter (or equivalent diameter, for a
rectangular stack or duct) is greater than 2.4 meters (7.8 ft), then you may
use 3-point sampling and locate the three points along the measurement line
exhibiting the highest average concentration during the stratification test, at
0.4, 1.0 and 2.0 meters from the stack or duct wall. If the gas stream is found
to be stratified because the 10.0 percent or 1.0 ppm criterion for a 3-point
test is not met, locate twelve traverse points for the test in accordance with
Table 1-1 or Table 1-2 of Method 1.
* * * * *
8.2.1 Calibration
Gas Verification. How must I verify the concentrations of my calibration gases?
Obtain a certificate from the gas manufacturer documenting the quality of the
gas. Confirm that the manufacturer certification is complete and current.
Ensure that your calibration gas certifications have not expired. This
documentation should be available on-site for inspection. To the extent
practicable, select a high-level gas concentration that will result in the
measured emissions being between 20 and 100 percent of the calibration span.
* * * * *
8.2.4
NO2 to NO Conversion Efficiency Test. Before or after
each field test, you must conduct an NO2 to NO
conversion efficiency test if your system converts NO2
to NO before analyzing for NOX. You may risk testing
multiple facilities before performing this test provided you pass this test at
the conclusion of the final facility test. A failed final conversion efficiency
test in this case will invalidate all tests performed subsequent to the test in
which the converter efficiency test was passed. Follow the procedures in
Section 8.2.4.1, or 8.2.4.2. If desired, the converter efficiency factor
derived from this test may be used to correct the test results for converter
efficiency if the NO2 fraction in the measured test gas
is known. Use Equation
7E-8 in Section 12.8
for this correction.
* * * * *
9.0 Quality Control * * *
|